[Sense]:Mixer Subsection - Agilent Technologies ESA-E Series Programmer's Manual

Esa spectrum analyzers
Hide thumbs Also See for ESA-E Series:
Table of Contents

Advertisement

[:SENSe]:HARMonics:SWEeptime?
Set the sweep time used for measuring each harmonic.
Factory Preset
and *RST:
200/resolution bandwidth (RBW)
Range:
10 ms to the maximum sweep time of the analyzer
Default Unit:
seconds
Remarks:
When set to 1 (On), the sweep time for measuring each
harmonic will be set to 200/RBW, where RBW is the
resolution bandwidth setting in Hz. When set to 0 (Off),
you can specify any sweep time within the limits of the
analyzer, with a minimum of 10 ms. This sweep time is
used only for the zero span measuring of the harmonics,
and not during the search for the fundamental. In this
case, the sweep time used will be the sweep time set
before the measurement began.
Front Panel
Access:
Meas Setup, ST/Harmonic
Turn On or Off Fundamental Zoom
[:SENSe]:HARMonics:ZOOM OFF|ON|0|1
[:SENSe]:HARMonics:ZOOM?
Turn on or off the fundamental signal zoom before the frequency count
occurs.
Factory Preset
and *RST:
On
Remarks:
When counter zoom is On, the measurement will zoom
in on the fundamental after the original marker search
and before the frequency count is executed. This
ensures the accuracy of the frequency count, and is
necessary when the starting resolution bandwidth is
very small or the starting span is very wide.
Front Panel
Access:
Meas Setup, Counter Zoom

[SENSe]:MIXer Subsection

Select External Mixer Band
[:SENSe]:MIXer:BAND K|A|Q|U|V|E|W|F|D|G|Y|J|USER
[:SENSe]:MIXer:BAND?
Chapter 5
Language Reference
SENSe Subsystem