Installation

Remove the Shipping Hardware

Shipping Hardware. All shipping hardware must be removed before
-CAUTION-
operating the instrument and reinstalled before repackaging the
instrument for shipment.
Pull open the microplate carrier access door.
1.
Use the supplied screwdriver to remove the two shipping bracket holders (PN
2.
1032190).
Remove the carrier shipping bracket (PN 1332048) by lifting it up to disengage
3.
the pin from the floor of the instrument.
Store the bracket and holders in case the instrument needs to be shipped.
4.
Remove the tape covering the cuvette port, if equipped.
5.

Select an Appropriate Location

Install the reader on a level, stable surface in an area where an operating temperature
between 18°C and 30°C (models with a touchscreen; "TS") or 18°C and 40°C (models
without a touchscreen; "NS") can be maintained.
The reader is sensitive to extreme environmental conditions. Avoid:
• Excessive humidity. Condensation directly on the sensitive electronic circuits can
cause the instrument to fail internal self-checks. The humidity must be in the range
of 10–85%, non-condensing.
• Excessive ambient light. Bright light may affect the reader's optics and readings,
reducing its linear range.
• Dust. Readings may be affected by extraneous particles in the microplate wells. A
clean work area is necessary to ensure accurate readings.
Agilent BioTek Epoch 2 Instructions for Use
1771011I Revision C
Page 17 of 51