Adjustments and Correction Constants

Source Spur Avoidance Tracking Adjustment

Source Spur Avoidance Tracking Adjustment
This adjustment optimizes tracking between the YO (YIG oscillator) and the cavity
oscillator when they are frequency offset to avoid spurs. Optimizing YO-cavity oscillator
tracking reduces potential phase-locked loop problems.
Analyzer warmup time: 30 minutes.
Required Equipment and Tools
Description
TORX screwdriver T-15
BNC alligator clip adapter
BNC-to-BNC cable
Antistatic wrist strap
Antistatic wrist strap cord
Static-control table mat and ground wire
Procedure
1. Remove the upper-rear corner bumpers and top cover, using a torx screwdriver.
2. Mate the adapter to the BNC cable and connect the BNC connector end to AUX INPUT
on the analyzer's rear panel. Connect the BNC center conductor alligator-clip to A11
TP10 (labeled φ ERR); the shield clip to A11 TP1 (GND) as shown in
Figure 3-24
Location of A11 Test Points and A3 CAV ADJ Adjustments
3. Press
Preset
Center
3-50
HP/Agilent Part Number
8120-1292
8120-1840
9300-1367
9300-0980
9300-0797
400
M/µ
Span
50
Figure
.
M/µ
3-24.
Chapter 3