Connect the Test Equipment
Connect the 16962A Logic Analyzer Pod to the Pulse Generator
16962A Logic Analyzer Service Guide
Connect Transition Time Converters (if required—see
1
page
25) to each of the four outputs of the pulse
generator: Channel 1 OUTPUT, Channel 1 OUTPUT,
Channel 2 OUTPUT, Channel 2 OUTPUT.
2
Connect the two SMA/Flying Lead test connectors (see
"Assemble the SMA/Flying Lead Test Connectors"
page 26) with 50 ohm terminators to the Transition Time
Converters at the pulse generator Channel 1 OUTPUT and
Channel 1 OUTPUT. (If Transition Time Converters are
not required, connect the SMA/Flying Lead test
connectors directly to the pulse generator outputs.)
Connect the two SMA/Flying Lead test connectors without
3
50 ohm terminators to the Transition Time Converters at
the pulse generator Channel 2 OUTPUT and Channel 2
OUTPUT. (If Transition Time Converters are not required,
connect the SMA/Flying Lead test connectors directly to
the pulse generator outputs.)
Connect one E5382A Flying Lead Probe Set to Pod 1 of
4
the 16962A logic analyzer. This probe set is used for data
inputs.
Testing Logic Analyzer Performance
3
on
35