The query performs the alignment and returns a zero if the alignment
is successful.
Before executing this command, connect a cable between front panel
NOTE
connector
spectrum analyzers except Agilent models E4401B and E4411B. Except
for the models indicated that do not require the cable, the alignment
will fail using command CAL:RF if the cable is not connected.
Front Panel
Access:

Select the Source State for Calibration

:CALibration:SOURce:STATe OFF|ON|0|1
:CALibration:SOURce:STATe?
Controls the state of the 50 MHz alignment signal.
The alignment signal is internally switched to the
NOTE
models E4401B and E4411B. For all other models, connect a cable
between front panel connector
before performing a calibration.
Factory Preset
and *RST:
Front Panel
Access:

Calibrate the Tracking Generator

:CALibration:TG
:CALibration:TG?
Performs an alignment of the tracking generator assembly.
The query performs the alignment and returns a zero if the alignment
is successful.
Chapter 5
and the
AMPTD REF OUT
System, Alignments, Align Now, RF
AMPTD REF OUT
Off
For Agilent ESA models E4401B and E4411B:
Input/Output (or Input), Amptd Ref (f=50 MHz) On Off
For all other Agilent ESA models:
Input/Output (or Input), Amptd Ref Out (f=50 MHz) On Off
CALibration Subsystem
connector for all Agilent ESA
INPUT
INPUT
and the
INPUT
Language Reference
for Agilent
connector
5-35