3
Testing Logic Analyzer Performance

To re-align a stray channel

48
(although it may be shifted left or right), and Eye Finder
must be able to place the blue bars in the eye.
If the blue bar for a particular bit does not appear in its eye
near the recommended starting position, then do the
following steps to realign the sampling position of the stray