PE Stud and Pole Clamp

1.
Remove nut securing earth wire.
2.
Remove nut securing PE Stud.
3.
Remove the three screws and remove the pole clamp assembly.
4.
Reassemble in reverse order.
(D) Label protective earth
A modified Knob Pole Clamp screw has been designed which incorporates a metal tip added to its
end. This amendment has increased the holding friction of the Pole Clamp assembly on the pole and
reduced the amount of tightening required to grip the pole.
The Pole Clamp Arm spares kit replaces parts of the Pole Clamp assembly to address bent or slipping
pole clamps. Note: There is no requirement to remove the V Clamp.
Item
Description
A
ASENA SP, KIT, PE STUD
B
ASENA SP, ASSY, POLE CLAMP
C
ASENA GW, KIT, FIXINGS(SCREWS,WASHERS,ETC)
D
LABEL PROTECTIVE EARTH
E
SPARES KIT POLE CLAMP ARM
F
POLE CLAMP SNAKE EYE DRIVER (not shown)
G
Spare pole clamp arm kit asena sp/gw
Alaris® GW Volumetric Pump
Spare Parts Replacement Procedures
V Clamp (remains fixed to the
pump case).
Arm clamp
Modified Knob Pole
Clamp Screw.
(C) Fixing screws (x3)
(A)
(A)
(B) Pole clamp assembly
Apply small amount of Castrol LMX
grease to this surface.
Apply Loctite 242 &
(E)
Loctite 7471 here.
Part Number
1000SP00467
1000SP00115
1000SP00489
1000LB00292
1000SP00589
1000ME01466
1000SP00589
48/77
(A) PE stud
Spring
Ball Bearing
Pivot Screw (Torque to 2Nm
using Driver)
(G)
1000SM00006 Issue 4
(A)