Changing the Air Filter

1. To make sure that the electrical components in the spectrometer are properly cooled, the air
filter must allow sufficient air into the instrument. Check the filter regularly and change it if it
is dirty.
You should not need to change the filter more than once every year if the environmental
conditions are as described in the Spectrometer Maintenance Checklist of this chapter.
Replacement filters are available; see Replacement Parts and Accessories later at the end of
this chapter, for the part number.
2. Locate the air filter on the right side of the instrument.
3. Depress the latch and lift out the plastic grid that holds the air filter in place.
4. Replace the air filter and then put the grid back into place.
Figure 9- 1 Air Filter
AAnalyst 200 User's Guide
9-11