Operation of the Instrument
Removing lamps
1. Do not unplug and remove lamps while they are switched on.
2. Touch the Lamp tab to display the Lamp page.
3. On the Lamp page, touch the Install Lamps button. The Install Lamps dialog is displayed.
4. Make sure that the lamp is switched off before you remove it. The On/Off box must not be checked.
5. If necessary, touch the respective On/Off box to switch off the lamp.
6. Touch OK.
7. Unplug and remove the lamp. For EDLs, remove the coding plug as well.
Tips for installing lamps
Use PerkinElmer lamps.
Do not touch the front window of the lamp: perspiration or other contamination can reduce
intensity of the radiation.
With EDLs, make sure that you use the correct coding plug and connect it to the receptacle on the
lamp holder containing the EDL.
If you use the wrong coding plug, you will cause the spectrometer to malfunction and may cause
irreparable damage to the lamp.
6-24