Removing the Burner Assembly from the Instrument
1. Extinguish the flame, turn off the gases, and power down the instrument. Let the burner
head cool. Open the burner door.
2. Disconnect the drain tube, unplug the drain interlock connector.
3. Locate the locking pin on the burner assembly and pull upward to unlock the lever.
Figure 9- 5 Burner side view
4. Push the lever forward to release the burner assembly.
5. Slide the assembly out and place the burner assembly on a bench.
6. The assembly is now ready for servicing.