Screen 8- 10 Parameters page, Calibration subpage.
Parameter entry is now complete on the Parameter page. At this point, the autosampler tray
should be filled with blank, standards, and samples. All of the parameter entries on the Analyze
page should be completed by now. Proceed to the next step.
Analyze Page
The results appear on the analyze screen in the order of the solutions analyzed. The most recent
results are displayed at the top of the screen.
Touch the Analyze tab to display the Analyze page.
Touch the Printer On/Off button if you wish the results to be printed.
Calibrate
Use this procedure to establish your calibration curve before analyzing samples. This procedure
will analyze your blank and standards only.
1. Touch the Calibrate button on the Analyze Sample screen. A green light indicates that it has
started.
2. Touch the Display Calibration button to view the standard calibration curve.
AAnalyst 200 User's Guide
8-19