Operation of the Instrument

Safety Checks

Before igniting the flame, always make sure that:
The exhaust venting system for the laboratory is switched on.
Burner chamber is correctly installed.
The burner head is correctly installed.
The burner end cap is secured.
The nebulizer is installed.
The correct gaskets are fitted in the burner system
The burner drain is connected to the system secured by clamp.
The drain trap is filled and is operating correctly.
The door of the atomizer compartment is in the closed position.
Remove all nearby flammable chemicals.
Fire extinguisher should be kept nearby.
Check for chemical residue build-up on the burner head slot.
When igniting nitrous oxide flame, warm up the burner head with air-acetylene flame for 15
minutes.
All safety interlocks are correctly closed.
If the interlocks are not closed, one or more of the following may be the cause:
The burner head, the nebulizer, or the drain system is not correctly installed.
The fuel or oxidant pressure is too low.
The liquid level in the drain trap is either too low or too high.
6-26