2. Clean the capillary assembly using a mild laboratory detergent solution, or a solvent such as
isopropanol, preferably in an ultrasonic bath.
3. Clean the nebulizer body and any other parts that appear dirty in a mild laboratory detergent
solution, or a solvent such as isopropanol, preferably in an ultrasonic bath.
4. Rinse all the cleaned components thoroughly with deionized water.
5. Inspect the o-rings (1 on the capillary assembly, and two on the nebulizer body) and clean
them if necessary in mild laboratory detergent solution, followed by a thorough rinsing with
deionized water.
If they look damaged, replace them.
6. After you have cleaned the components, reassemble the nebulizer.