Figure 9- 9 Exploded view of the burner chamber assembly
To install the burner in the spectrometer
1. Guide the burner assembly into the tracks of the atomizer compartment.
2. Push the burner assembly in so that the rear gas fittings line up with the gas connectors at the
back of the atomizer compartment. Push the burner assembly fully in until it clicks into place.
Pull the lever forward and insert the locking pin to lock the burner assembly in place.
3. Locate the drain vessel. The drain tube and interlock cable will be attached to the drain vessel
retainer cap.
4. Place the drain vessel in full view on the floor in front of the instrument. Never place the drain
vessel inside a cabinet.
AAnalyst 200 User's Guide
9-29