On Parameters page, Spectrometer subpage:
1. Enter 1 second for the Integration Time.
2. Enter 3 for the number of Replicates.
3. Enter 5 seconds for the Read Delay.
4. Enter the required Print parameters.
5. Select Manual Data Display for Sample Handling.
Screen 8- 1 Parameters page, Spectrometer subpage.
On Parameters page, Calibration subpage:
Select the Calibration Equation.
1. For this example select Linear Through Zero. Enter mg/L for Units.
On Parameters page, Calibration subpage, Standard Concentration section:
1. Enter the concentration of the first standard for the element that you are determining (e.g., 0.5
mg/L for copper).
2. Enter the concentration of Standard 2 for the element you are determining (e.g., 1.0 mg/L for
copper).
3. Enter the concentration of Standard 3 for the element you are determining (e.g., 1.5 mg/L for
copper).
For the above example we will only be calibrating with three standards.
AAnalyst 200 User's Guide
8-7