Maintenance
296
Item
2
Spray Chamber
3
PVC Tubing 1.5 mm i.d. (Part No.
09985735)
4
Coupler (Part No. 09920186)

General System Maintenance

Cleaning the Instrument

Exterior surfaces may be cleaned with a soft cloth, dampened with a mild detergent
and water solution. Do not use abrasive cleaners or solvents.

Air Filters

To ensure that the electrical components in the spectrometer and the RF generator
are properly cooled, check the air filters regularly to see if they are clogged or dirty.
There are four air filters on the right and left sides of the instrument as shown in
Figure 5-41.

Changing the Air Filters

The air filters are located inside the snap-on plastic grid. Refer to the following
figure.
1. Shut down and switch off the spectrometer.
2. Carefully pull off the plastic grid on the ventilator.
3. Remove the filter and insert a new one (Part No. 09995098).
4. Carefully push the plastic grid onto the ventilator.
Description
Item
Description
6
Teflon Tubing 1 mm i.d. (Part No.
09985708)
7
Pump Tubing 1.14 mm i.d.
8
Peristaltic Pump