Analyzing Samples
Touch the Analyze tab to display the Analyze page.
Touch the Printer button if you wish the results to be printed.
NOTE: Check printer function, the printer already maybe on.
1. Aspirate the blank
2. Touch Analyze Blank button.
3. Aspirate Standard 1, and then touch the Analyze Standard button.
NOTE: After a standard is analyzed, the next standard concentration is automatically incremented in the
standard entry-field box. You have the option of using the automatically filled entry or you may
select a different standard concentration to be analyzed by using the standard touch entry-field
box . The dialog displays a list of standard concentrations that you can choose from.
4. Aspirate Standard 2, then touch the Analyze Standard button.
5. Aspirate Standard 3, then touch the Analyze Standard button.
6. To view the calibration curve, touch the Display Calibration button.
7.
Aspirate sample 1, then touch the Analyze Sample button.
NOTE: After a sample is analyzed, the next sample ID is automatically incremented in the sample entry-
field box. You may enter a sample number by touching the sample-entry field box to use the
keypad.
8. Aspirate sample 2, then touch the Analyze Sample button.
8-14
Screen 8- 8 Analyze page