PerkinElmer aanalyst 200 User Manual page 131

Table of Contents

Advertisement

Adding water to the Drain Float Assembly
Enough water must be present in the drain system to fill the drain loop and to activate the drain float
assembly. If not, carry out the procedure below:
1. Loosen the drain clamp and remove the drain tubing from the drain outlet on the burner end cap (see
Figure 6- 8).
2. Pour about 250 mL of water into the drain tubing.
3. Reconnect the drain tubing to the burner end cap. Replace the drain clamp and tighten securely.
Installation of the burner drain is complete.
Installing the Lamp Compartment Door
Locate the lamp compartment cover (if it is not already on the instrument). Install the cover by following
the procedures below.
1. Hold the door in place next to the lamp compartment as shown below.
2. Gently push the upper part of the door on to the upper pin and maneuver the lower part of the door
over the screw hole. Tighten the securing screw with a screwdriver.
3. Close the door.
AAnalyst 200 User's Guide
6-17

Advertisement

Table of Contents
loading

Table of Contents