Analyzing Samples
When you are satisfied with your calibration curve you may proceed to the next step.
Analyze Samples
Use this step to analyze samples only.
1. Touch the Analyze Samples button to begin analyzing samples. A green light indicates that it
has started.
2. Touch the Display Calibration button to view the calibration curve at anytime during the
run.
3. Proceed to the next step.
NOTE: To stop analyzing samples at anytime, touch the Analyze Sample button to turn the function off.
Analyze All
This step will start analyzing your entire run (blank, standards, and samples).
1. Touch the
2. Touch the Display Calibration button to view the standard calibration curve at anytime
during your run.
NOTE: To stop the run, touch the Analyze All button to turn the function off. If you were in the middle of
running your samples and would like to continue the run, you must change the parameter of First
Sample Location on the Analyze Samples page to the desired sample location then touch the
Analyze Samples button.
8-20
Screen 8- 11 Display Calibration screen.
button to begin your run. A green light indicate that it is has started.
Analyze All