Touch Screens
For an uncoded lamp, you must complete the following entries: Element, Lamp Type, and
current. The Wavelength and Slit Width are automatically filled in after entering the element
symbol.
Coded Lamp
Single-Element Coded Lamp
1. If you installed a single-element coded lamp, the analyte appears automatically in the
element entry-field box.
2. To change entries of other recommended parameters (Wavelength, Slit, Signal), touch the
entry-field box to display dialogs. Touch the Install Lamp button to make changes to the
recommended parameters in the Install Lamp dialog.
3. When lamp parameters are properly set up, touch the Setup Instrument button.
Multi-Element Coded Lamp
1. If you installed a multi-element coded lamp, touch the Elements entry-field box to
display the Element dialog. Select the required analyte element from the list in the dialog
and the recommended parameters for that element are set up.
2. To change entries of other recommended parameters (Wavelength, Slit Width, Signal);
touch the entry-field box to display dialogs. Touch the Install Lamp button to make
changes to the recommended parameters in the Install Lamp dialog.
3. When lamp parameters are properly set up, touch the Setup Instrument button.
Uncoded Lamp
If you installed a single-element or multi-element uncoded lamp follow the steps below.
1. From the Lamp page, touch the Install Lamp button. The Install Lamp dialog appears
and you must manually input the required entries for each of the parameters. To input
entries, touch on the entry-field box and a dialog or keypad will display.
2. After all entries are made for each of the parameters, touch OK to return to the Lamp
page.
3. On the Lamp page, touch the Elements entry-field box to select an element.
4. When lamp parameters are properly set up, touch Setup Instrument button.
7-10