4. Lower the nebulizer clamp until it presses against the barrel and flange of the nebulizer (shaded area).
Ensure that the nebulizer is pushed in all the way and clamped in correctly. Warning, the nebulizer
clamp must be lowered and locked OVER the nebulizer flange. Detail B shows the clamp lowered
and locked correctly over the nebulizer flange.
5. Next you'll connect the nebulizer oxidant hose to the nebulizer side arm. Remove the plastic nut from
the nebulizer side arm. Slide the nut on the end of the hose hanging free on the burner base. The hose
has white tape on it printed with the word "NEB."
6. Slide the NEB oxidant tubing onto the ridges on the nebulizer side arm.
7. Using the Apiezon grease, lightly lubricate the OUTSIDE of the tubing where the nut will screw onto
the nebulizer ridges.
8. Finally, slide the plastic nut up the black tubing and gently thread it over the ridges on the side arm.
Tighten the plastic nut just finger tight. The nebulizer is now assembled in place and secure.
Figure 6- 2 Installing nebulizer onto the burner end
AAnalyst 200 User's Guide
6-7