Connecting the Electrical Supply

A Review of the Electrical Specifications
The electrical supply voltage must be between: 100-230 VAC ± 10%
The electrical supply voltage should be free of line transients in excess of 50V peak. If the
supply voltage produces large AC line voltage fluctuations, install a voltage regulator between the
AC outlet and the instrument.
The spectrometer requires either a 100/120 VAC standard 15 amp grounded outlet, or a 230 VAC
outlet. In addition, the instrument consumes 300 volt amps (maximum) when configured for
background correction.
The AAnalyst 200 requires a grounded AC outlet providing voltage corresponding to the voltage
range marked on the instrument (100-230 VAC).
Do not remove or alter the ground prong on a three-prong power
plug.
WARNING
AAnalyst 200 User's Guide
4-19