Cleaning (or flushing) the Drain System
It is recommended that the drain system be flushed thoroughly with water at the end of each
working day to remove caustic, corrosive, or organic waste materials that could otherwise
damage the burner chamber or drain tubing.
To flush the system, remove the burner head then slowly pour about 500 mL of water into the
burner chamber. Allow the water to drain into the drain vessel.
Replacing the Drain Tubing
If the drain tubing (P/N 0250-7987) is damaged or extremely dirty and needs to be replaced, refer
to the next figure and proceed as described next.
Dispose of drain vessel waste in accordance with the regulations applicable to
your locality, state, and/or country.
WARNING
Figure 9- 11 Replacing the drain tubing
AAnalyst 200 User's Guide