Analyzing Samples

Manual Analyses: Data Display MHS-Technique

If you have selected MHS technique from the Startup dialog, continue as described in this
chapter.
Setting Up MHS-15 Analyses
1. Read the safety information in the handbooks for your system before you operate the
spectrometer.
2. Make sure that the analyzer assembly and quartz cell are correctly installed. See the
handbooks for your system. Switch on the fume ventilation system for the work area.
3. Switch on the AA200 instrument. See Chapter 6, Operation of the Instrument.
4. Select the MHS technique from the startup screen or from the tools menu → startup.
5. Select or fill the required touch entry fields on the Parameters page, Spectrometer subpage.
6. Install and align the lamps that are required. See chapter 6, for "Aligning Lamps".
7. Follow the procedures for setting up Flame Analyses including setting up and adjusting the
burner and igniting the flame in chapter 6.
8. Once you are ready to perform the analysis, see the procedures for Manual Flame Analyses in
this chapter.
NOTE: To initiate the read, touch the Analyze Samples button on the Analyze page. Wait for the
specified BOC time to elapse (typically 2 seconds), then press the plunger down on the MHS-15.
8-22
Screen 8- 13 Parameters page, Spectrometer subpage.