Maintenance
Replacing the E-Box
Using excessive force and not aligning the module on the track can damage
CAUTION
the E-Box.
1. Holding on to the E-Box with the handles, position the E-Box to align on the module
track.
2. Use the handles to gently push the E-Box until it is secured in place.
3. Screw the locking screw to fully secure the E-Box.
4. Reconnect the necessary cables to the ports on the module.
5. Reattach the compartment door.
6. Connect the AC line cord to the E-Box and power up the instrument.
7. Close the compartment door.
9-48