Adding Water to the Drain Float Assembly
Enough water must be present in the drain system to fill the drain loop and to activate the drain
float assembly. If not, carry out the following procedure:
1. Loosen the drain clamp and remove the drain tubing from the drain outlet on the burner
end cap.
2. Pour about 250 mL of water into the drain tubing.
3. Carefully reconnect the drain tubing to the burner end cap. Replace the drain clamp and
tighten securely. Installation of the burner drain is complete.
Figure 9- 12 The drain trap assembly.
AAnalyst 200 User's Guide
9-35