General Laboratory Requirements
Operating conditions
Warning: Explosive Atmosphere
This instrument is not designed for operation in an explosive
atmosphere.
WARNING
Recommended operating conditions:
•
Indoors.
•
The location must be free of smoke, dust, corrosive fumes, direct sunlight and excessive
vibration.
•
Temperature: +15 °C to +35 °C (+59 °F to +95 °F). Maximum rate of change not exceeding
3 °C (5 °F) per hour.
•
Relative humidity: 20% to 80%, without condensation.
•
Altitude: in the range 0 m to 2,000 m (sea level to 6562 feet).
Safe operating conditions:
•
Indoor.
•
Temperature: +5 °C to +40 °C (+41 °F to +104 °F).
•
Relative humidity: 20% to 80%, without condensation.
•
Altitude: in the range 0 m to 2,000 m (sea level to 6562 feet).
Storage conditions:
•
Ambient temperature: –20 °C to +60 °C (–4 °F to +140 °F).
•
Relative humidity 20% to 80%, without condensation.
•
Altitude in the range 0 m to 12,000 m (sea level to 39,370 feet).
AAnalyst 200 User's Guide
3-5