4. Screw the connector nut onto the nebulizer side arm. Tighten the connector nut so that the
side-arm threads are covered by the nut.
Excessive tightening could result in damage to the nebulizer when you
CAUTION
remove the nut.
5. Apply a small amount of Apiezon grease onto the o-rings (P/N 0990-2015). The nebulizer
will then slide easily into the end cap when you perform steps 5 and 6, below.
6. Lift the end-cap spring clamp and insert the nebulizer in the center of the end cap. Then
release the spring clamp while you make sure the spring clamp slips over the nebulizer
flange.
Use care installing the nebulizer as the impact bead is quite fragile.
CAUTION
7. Rock the nebulizer slightly to make sure it is set correctly in the end cap. If excessive
movement doesn't occur, it is set properly.
8. The nebulizer installation is complete.
AAnalyst 200 User's Guide