Touch Screens
8. The Microsoft ActiveSync window appears and displays the status of the connections
between the E-Box and the computer.
NOTE
: A green icon
connection.
Screen 7- 32 The Microsoft ActiveSync window shows the connection.
9. Connection is established.
Downloading
Once the connection is established between the computer and the E-Box, continue with the
procedure below for downloading.
1. Insert the QuickStart CD (0993-6637) into the computer and use the browser to find
Update.bat file.
2. Double click on the Update.bat.
3. The command prompt window appears and the download of the software/ firmware to the E-
Box begins.
4. A successful download will display 'Download is Completed' on the command prompt.
5. Reboot the AAnalyst 200 instrument by powering it down and back up.