Touch Screens
Replicates
Buttons on Analyze Samples page
Analyze All
Calibrate
Analyze Samples
Display Calibration
Printer On/Off
Parameters on Analyze Samples page
Autosampler Tray
First Sample
Location
Last Sample
Location
Recalibration
Frequency
7-26
referred to as the coefficient of variation.
This displays the replicate information for each
replicate analyzed.
By touching this button, the system will start an
automated analysis of the blank, standards, and
samples. A green light appears on the button when
selected.
Touch this button to start the calibration sequence. The
system uses the blank and standards defined in the
method.
To analyze samples, touch this button. A green light
appears on the button when selected. The system will
analyze sample range specified in the Sample
Locations entry-field box.
To display the calibration curve window, touch this
button.
By default the print function is On and the results will
print when the printer queue is full. The green light
indicates that it is On. To turn print function Off, touch
Printer On/Off button. This will print any results in
the printer queue.
This shows the name of the sample tray currently
selected. To select a different tray, touch this entry-
field for tray dialog.
This shows the starting position of the sample located
on the autosampler tray. To change the start sample
location, touch this entry-field for a number-pad entry
dialog.
This shows the terminating position of the sample
located on the autosampler tray. To change the
terminating sample location, touch this entry-field for a
number-pad entry dialog.
The system analyzes the calibration blank and all the
calibration standards (or reslope if selected and
defined), then creates a completely new calibration
curve. To set the frequency, touch this entry-field for
the number-pad dialog.