Continuous Display
If you selected Manual Continuous Graphic Display for Sample Handling on Spectrometer
subpage, continue as described in this section.
Enter parameters on Lamp page
1. If not already done, install and align the required lamp. For more information see Chapter 6,
Operation of the Instrument.
2. Enter lamp parameters. For more information on how to enter analytical parameters refer to
Chapter 7, Touch Screen.
On Flame page
1. Install burner head see Chapter 6, Operation of the Instrument.
2. Adjust the burner height see Chapter 6, Operation of the Instrument.
3. Adjust fuel/oxidant flows.
4. Check interlocks see chapter 7, Touch Screen.
5. Ignite flame see Chapter 6, Operation of the Instrument.
6. Adjust the nebulizer see Chapter 9, Maintenance.
7. Optimize the burner system see Chapter 6, Operation of the Instrument.
On Parameter page, Spectrometer subpage:
1. Enter 1 second for the Integration.
NOTE: The default for Replicate in continuous mode is 1.
2. Enter 0 seconds for the Read Delay. The default value is zero.
3. Enter the required Print parameters.
4. Select the information you would like to have printed together with the analytical results.
5. Select Manual Continuous Graphic Display for Sample Handling.