Touch Screens
Buttons on Analyze Samples page
Analyze Blank
Analyze Standard
Analyze Sample
Display Calibration
Printer On/Off
Items on Analyze Samples page
Standard
Concentration Box
Sample Number Box
ID Number Box
Dilution Box
7-22
To analyze the blank defined in the method, touch this
button. A green light appears on this button when selected.
To analyze the standard indicated in the standard
concentration box, touch this button. A green light appears
on the button when selected. The system analyzes the
standard and creates a calibration curve. Standards can be
reanalyzed by selecting the standard from the Standard
Concentration Box.
To analyze the sample indicated by the sample number entry
field, touch this button. A green light appears on the button
when selected. Samples can be reanalyzed by selecting the
sample from the Sample Number Box.
To display the calibration curve window, touch this button.
By default the print function is On and the results will print
when the printer queue is full. The green light indicates that
it is On. To turn print function Off, touch Printer On/Off
button.
You may print the results without waiting for the printer
queue to fill up by touching the Printer On/Off button.
Touch this entry-field box to view the standard concentration
dialog. Use the dialog to select the standard concentration
you want to analyze. A calibration curve is created after
blank and selected standard are run.
This entry field shows the sample No. of the sample you are
currently analyzing. If you wish to analyze a different
sample, touch this entry-field box to view the sample
number dialog.
This entry field allows you to enter the ID number for the
sample.
This entry field allows you to enter the sample dilution.