Analyzing Samples
5. Select the information you would like to have printed together with the analytical results.
6. Select Autosampler for Sample Handling.
Screen 8- 9 Parameters page, Spectrometer subpage.
On Parameters page, Calibration subpage:
1. Select the Calibration Equation.
2. For this example select Linear Through Zero. Enter mg/L for Units
On Parameters page, Calibration subpage, Standard Concentrations section:
1. Enter the concentration of the first standard for the element that you are determining (e.g., 0.5
mg/L for copper).
2. Enter the concentration of Standard 2 for the element you are determining (e.g., 1.0 mg/L for
copper).
3. Enter the concentration of Standard 3 for the element you are determining (e.g., 1.5 mg/L for
copper).
For the above example we will only be calibrating with three standards.
•
Enter the concentration of the Reslope standard (e.g., 5 mg/L for copper).
8-18