Maintenance
5. (Optional) To completely remove the burner end cap, disconnect the auxiliary oxidant hose
and the fuel hose from the burner end cap using the "double wrench" method (one wrench to
hold the nebulizer fitting and the other to turn the compression fitting).
6. (Optional) Disconnect the drain tube and clamp (if they are in place).
7. (Optional) Unscrew the four knurled screws holding the end cap to the burner chamber and
carefully remove the end cap.
8. Remove the flow spoiler from inside the burner chamber
9-26
Figure 9- 8 Removing the burner end cap.