PerkinElmer aanalyst 200 User Manual page 282

Table of Contents

Advertisement

Maintenance
Figure 9- 13.
Then, while holding the nebulizer assembly in one hand, rock the gas supply tubing
with the other hand while gently coaxing the tubing from the barbs at the end of the
nebulizer side arm.
5. Remove the impact bead and, if using the impact bead spacer, remove the spacer as well.
Handle with care. (See the next page for instructions.)
6. Unscrew the capillary tubing.
7. Figure 9- 13 and loosen and then remove the adjustment nut (P/N N037-1492) and locking
ring.
8. Remove the capillary assembly from the nebulizer body.
Cleaning
1. While holding the capillary assembly in one hand, take one of the cleaning wires (P/N 0303-
0135) supplied.
2. Push it through the capillary assembly from its wider (open) end.
NOTE: It may take several attempts to pass the wire through the nebulizer.
9-38
Refer to

Advertisement

Table of Contents
loading

Table of Contents