5. Touch the On/Off check box. If necessary, change the lamp current. Touch OK.
6. Touch the Element entry field box to select the desired element. Touch OK.
7. Touch the Setup Instrument button on the Lamp page of the touch screen. The recommended
parameters for that element are automatically set up.
The system automatically switches the lamp on, sets up the optical path (prism and grating),
sets and peaks the wavelength, sets the slit, and performs an Automatic Gain Control (AGC)
and Automatic Intensity Control (AIC).
8. Turn the two alignment controls alternately and in small increments while watching the AA bar
graph and the numerical energy display. Align the lamp to obtain maximum energy.
Uncoded multi-element Lumina lamp
1. Follow steps 1-5 of the coded single element Lumina lamp.
2. Touch the Element entry field box on the Lamp page to display the Select Element dialog. Select
the desired elements. Use the Clear button to remove unwanted element from the selected list.
Touch OK.
3. Touch the Setup Instrument button on the Lamp page of the touch screen. The recommended
parameters for that element are automatically set up.
The system automatically switches the lamp on, sets up the optical path (prism and grating),
sets and peaks the wavelength, sets the slit, and performs an Automatic Gain Control (AGC)
and Automatic Intensity Control (AIC).
4. Turn the two alignment controls alternately and in small increments while watching the AA bar
graph and the numerical energy readout on Lamp page. Align the lamp to obtain maximum energy.
Turret coded Lumina lamp (mount)
1. Start up the spectrometer, but do not ignite the flame.
2. If the desired element already appears in the Element entry field box, then just touch the Setup
Instrument button.
3. If you like to select a different element, touch the Element entry field box on the Lamp page to
display the Element dialog. Select the desired element and touch OK.
The system automatically switches the lamp on, sets up the turret and the optical path (prism
and grating), sets and peaks the wavelength, sets the slit, and performs an Automatic Gain
Control (AGC) and Automatic Intensity Control (AIC).
AAnalyst 200 User's Guide