Operation of the Instrument
Assembling the Quick-Change Burner Mount Assembly
Before the burner assembly can be put into the instrument, a few items need to be added or checked. The
following sections give full procedures to do so.
Installing the Nebulizer on the Burner
Refer to the figure below to install the nebulizer onto the burner end cap.
Flow Spoiler
O-Rings
Figure 6- 1 Installing the nebulizer onto the burner end cap.
1. Feed at least 6-8 inches (152-203 mm) of capillary tubing onto the nebulizer inlet.
2. Lightly lubricate the nebulizer o-rings with the Apiezon grease supplied in the nebulizer kit.
3. Locate the nebulizer clamp (Detail A) and slide it up while inserting the nebulizer into the burner end
cap. Be sure to orient the nebulizer side arm so that it will lie in the slot provided. (A gentle twisting
and pushing motion might be necessary to fully seat the nebulizer in the burner end cap.)
6-6
Chamber
Impact Bead
Structure
Spacer
Nebulizer Body
Nebulizer Side
Arm
Burner
Head
Burner
(Mixing)
End Cap
Drain