Analyzing Samples
Enter the concentration of the Reslope standard (e.g., 1.0 mg/L for copper).
Parameter entry is now complete. Proceed to the next step.
Screen 8- 2 Parameters page, Calibration subpage.
Analyze samples
As measurements are made, the results appear on the analyze screen. You may view the
calibration curve of the analyses at anytime by touching the display calibration button.
Touch the Analyze tab to display the Analyze page.
Touch the Printer On/Off button if you wish the results to be printed.
1. Run the blank
Aspirate the blank solution.
Touch the Analyze Blank button.
2. Run the standards
Aspirate Standard 1, then touch the Analyze Standard button.
Aspirate Standard 2, then touch the Analyze Standard button.
Aspirate Standard 3, then touch the Analyze Standard button.