Analyzing Samples
Table 8- 1 Suggested Standards
Element
Ag
Cu
Mg
Mn
Pb
3. For samples 1, 2 and 3, use Standards 1, 2, and 3, respectively, in this example.
Data Display
If you selected Manual Data Display for Sample Handling on Spectrometer subpage, continue as
described in this section.
Enter parameters on Lamp page
1. If not already done, install and align the required lamp. For more information see Chapter 6,
Operation of the Instrument.
2. Enter lamp parameters. For more information on how to enter analytical parameters refer to
Chapter 7, Touch Screen.
On Flame page
1. Install burner head see Chapter 6, Operation of the Instrument.
2. Adjust the burner height see Chapter 6, Operation of the Instrument.
3. Adjust fuel/oxidant flows.
4. Check interlocks see Chapter 7, Touch Screen.
5. Ignite flame see Chapter 6, Operation of the Instrument.
6. Adjust the nebulizer see Chapter 9, Maintenance.
7. Optimize the burner system see Chapter 6, Operation of the Instrument.
8-6
Standard 1
Standard 2
(mg/L)
(mg/L)
0.2
0.5
0.5
1
0.05
0.1
0.2
0.4
2
5
Standard 3
Reslope
(mg/L)
(mg/L)
1
0.5
1.5
0.2
0.1
0.6
0.4
10
1
5