Analyzing Samples
On Parameters page, Calibration subpage:
Select the Calibration Equation.
1. For this example select Linear Through Zero. Enter mg/L for Units.
On Parameters page, Calibration subpage, Standard Concentrations section:
1. Enter the concentration of the first standard for the element that you are determining (e.g., 0.5
for copper).
2. Enter the concentration of Standard 2 for the element you are determining (e.g., 1.0 for
copper).
3. Enter the concentration of Standard 3 for the element you are determining (e.g., 1.5 for
copper).
For the above example we will only be calibrating with 3 standards.
8-12
Screen 8- 5 Parameter page, Spectrometer subpage.