Data Flow (Remote Operation - Keithley 2750 User Manual

Multimeter/switch system
Hide thumbs Also See for 2750:
Table of Contents

Advertisement

Model 2750 Multimeter/Switch System User's Manual

Data flow (remote operation)

Remote operation can be used with triggering configured to perform a specified number of
measurements and then stop. The various read commands (SENS:DATA?, FETCh?,
READ?, MEAS?, CALC2:DATA?, TRACe:DATA?, and CALC1:DATA?) return the data
array(s) acquired during the measurement cycle. Data flow for this triggering configura-
tion is summarized by the block diagram shown in
for the following discussion.
Figure D-4
Data flow for remote operation
SENSe
Measurement, Filter, Rel,
and Ratio or Ch Avg
TRAC:CLE
INIT:CONT OFF
TRIG:COUN 1
INIT
Sample
Buffer
[SENS[1]]:DATA[:LAT]?
[SENS[1]]:DATA:FRESh?
CALC[1]:DATA[:LAT]?
CALC[1]:DATA:FRESh?
Signal Processing Sequence and Data Flow
Figure
D-4. Refer to this block diagram
CALC1
Math
(mX+B, Percent
or 1/X)
CALC3
Limit Tests
CALC3:LIM1:FAIL?
CALC3:LIM2:FAIL?
FETCh?
READ?
MEAS?
TRACe
Data Store
CALC
Min, Max, Sdev
Mean, Pk-Pk
CALC2:IMM?
CALC2:IMM
CALC2:DATA?
TRACe:DATA?
D-7