6
Rotate the level so that it faces away from you and the screw on the back of the level is on the right side.
Ensure that the level is positioned in the same spot as
completely settle. Without moving the level, gently press the HOLD button until an upside down [ 4 ]
appears on the display. At this time, all of the horizontal calibration settings are done.
7
Stand the level on its end so the display is facing you and the text on the face of the level can be read
from bottom to top. Mark the position of the level so you can place it in the same spot when you move
it during this vertical calibration process. Wait 10 seconds for the level to completely settle. Without
moving the level, gently press the HOLD button until a [ 5 ] appears on the display.
Scalar i6000 Installation Guide
Step
3. Wait 10 seconds for the level to
341