3
Place the level so that the display is facing you and lettering on the face of the level is right-side up.
Mark the position of the level so you can place the level back in the same spot when you move it during
this calibration process. Wait 10 seconds for the level to completely settle. Without moving the level,
gently press the HOLD button until [ 1 ] appears on the display.
4
Rotate the level so that it faces away from you and the screw on the back of the level is on the left side.
Ensure that the level is positioned in the same spot as
completely settle. Without moving the level, gently press the HOLD button until [ 2 ] appears on the
display.
5
Roll the level towards you so the display is facing you but the lettering on the face of the level is upside
down. Ensure that the level is positioned in the same spot as
completely settle. Without moving the level, gently press the HOLD button until an upside down [ 3 ]
appears on the display.
340
Step
3. Wait 10 seconds for the level to
Step
3. Wait 10 seconds for the level to
Testing and Calibrating the Digital Level