Internally Mounted Disk Drive - Keysight Technologies X Series Service Manual

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Assembly Replacement Procedures
Disk Drive

Internally Mounted Disk Drive

Drive Removal and Replacement
1. Turn the instrument off and remove the AC power cord.
2. Before replacing the A5 disk drive it is highly recommended that the
3. Refer to
546
factory calibration data be backed up to an external drive if at all possible.
If this data is not backed up prior to replacing the disk drive all instrument
adjustments and performance verification tests will need to be run after
the drive is replaced.
For information on how to backup this data see the
Backup and Restore"
section in
If this data cannot be backed up because the instrument will not fully
boot, go ahead and change the A5 Disk Drive and perform all adjustments
listed in
Chapter 16, "Post-Repair Procedures."
Figure
15-63. Locate and remove the existing CPU assembly from
the instrument by removing the 6 rear panel screws (1). The CPU
assembly can be removed from the chassis by pulling straight out the
back. Use the two ejectors to pull the CPU assembly out from the chassis.
"Calibration Data
Chapter 16
of this manual.
for the A5 assembly.
Keysight N9010A EXA Service Guide

Advertisement

Table of Contents
loading

Table of Contents