Post-Repair Procedures
Post-Repair Procedures
S/W Revision Greater Than or Equal To A.02.00
1. Connect a USB keyboard and mouse to two of the instrument rear panel
2. Connect a USB FLASH drive to one of the front panel USB ports.
3. Press System, Alignments, Backup or Restore Align Data.
4. The Alignment Data Wizard window as shown in
Figure 16-2
Alignment Data Wizard
5. Follow the on-screen instruction to backup the calibration data to the USB
6. If the CPU is a PC6 or PC7 and has the SD memory card installed, also
600
USB ports
FLASH drive.
create a backup on the SD memory card.
Figure 16-2
will appear.
Keysight N9010A EXA Service Guide