Assembly Replacement Procedures
Front Frame Assembly
5. Refer to
Figure 15-92
Inverter Board Removal
6. To remove the A1A4 LCD Inverter board (1), remove the two screws (3)
7. To remove the A1A3 LCD (4), flip the assembly over and remove the four
Keysight N9010A EXA Service Guide
Figure
15-92. Disconnect the backlight wires (2) from the
inverter board.
securing the inverter board to the display bracket (5).
screws securing the LCD to the display bracket (5).