Allows the YTF to be bypassed, improving amplitude accuracy.
Assure start frequency is 3.6 GHz or greater. Press Amplitude, uW Path
Control, uW Preselector Bypass.
When the bypass switches Switch 1 and Switch 2 change state you will hear a
click. Depending on the signal source connected, you may see many "signals"
appear on screen because without pre-filtering, the analyzer will display image
and multiple responses. The real input signal should not change amplitude
when you switch the uW Preselector Bypass on and off.