Instrument Messages
-221 Settings Conflict Errors
-221
Settings conflict; Settings
conflict; Mask unavailable
for current Span. Increase
to display mask.
-221
Settings conflict; Settings
conflict; Pre-trigger is
insufficient for demod.
Decrease Trig Delay.
-221
Settings conflict; Signal
Track is not available when
Freq Scale=Log
-221
Settings conflict; Signal
Track is not available with
Continuous Peak
-221
Settings conflict; Signal
Track is not available with
Segmented Sweep
-221
Settings conflict; Signal
Track is only available in
Swept SA measurement
-221
Settings conflict; Signal
Track is turned off when
Zero Span is selected
-221
Settings conflict; Span is
not available when Segmented
Sweep is on
-221
Settings Conflict; Span
limited to XXX
-221
Settings conflict; Span Zoom
is not available in Zero
Span
-221
Settings conflict; Span Zoom
not available when Frequency
Scale Type = Log
-221
Settings conflict; Span Zoom
not available when Segmented
Sweep is on
-221
Settings conflict; Step keys
are not available to modify
this function
-221
Settings conflict; Sweep
Setup only available in
swept measurements
Keysight N9010A EXA Service Guide
The current span setting is either narrower than the mask width or
so wide that there are too few display points to allow the mask to
be drawn. Increase or decrease the span to display the mask.
The signal tracking feature cannot be used when you have selected
a log scale for the frequency axis.
The signal tracking feature cannot be used while you are also using
the continuous peak function.
The signal tracking feature cannot be used when you are using the
segmented sweep capability.
The signal track functionality can be used when making a swept SA
measurement. It is not available in the SA measurement when you
are using FFT sweeps.
Signal Track is not available when you have selected Zero Span. So
if Zero Span is entered while in Signal Track is On, Signal Track is
turned off.
Segmented Sweep uses multiple sweeps, each with its own span
setting. You cannot set the span when segmented sweep is
selected because the function does not know which span to
change.
Span Zoom does not work with a time domain x-axis. You must
select a span greater then 0 Hz.
The Span Zoom feature cannot be used when the X-axis scale is
logarithmic frequency.
Segmented Sweep uses multiple sweeps so it does not know what
sweep to zoom in on. Also, it cannot zoom a span over a sweep
boundary.
You should select a specific value for this function. So using the
Up/Down step keys to scroll through values is not allowed.
The current measurement uses FFT mode and so does not use the
Sweep Setup menu
101