Keysight Technologies X Series Service Manual page 567

Exa signal analyzer, 9 khz - 3.6/7/13.6/26.5/32/44 ghz
Hide thumbs Also See for X Series:
Table of Contents

Advertisement

Assembly Replacement Procedures
Motherboard Assembly
6. Replace the Front Frame Assembly. Refer to the
7. Replace the Midplane assembly. Refer to the
8. Replace the L.O. Synthesizer assembly. Refer to the
9. Replace the Front End Control assembly. Refer to the
10.Replace the AIF/DIF assembly. Refer to the
11.Replace the power supply assembly. Refer to the
12.Replace the CPU assembly. Refer to the
13.Replace the reference assembly. Refer to the
14.Replace the top brace and reference bracket. Refer to the
15.Replace the rear panel. Refer to the
16.Replace the instrument outer case. Refer to the
Keysight N9010A EXA Service Guide
replacement procedure.
replacement procedure.
Assembly
replacement procedure.
Assembly
replacement procedure.
(DIF) Assembly
replacement procedure.
replacement procedure.
procedure.
replacement procedure.
Reference Bracket
replacement procedure.
replacement procedure.
Front Frame Assembly
Midplane Assembly
L.O. Synthesizer
Front End Control
Analog I.F. (AIF)/Digital I.F.
Power Supply Assembly
CPU Assembly
replacement
Reference Assembly
Rear Panel
replacement procedure.
Instrument Outer Case
Top Brace and
567