Instrument Messages
-221 Settings Conflict Errors
-221
Settings conflict; Marker
Function is not available
for a Fixed marker
-221
Settings conflict; Marker
type must be delta
-221
Settings conflict; Marker->
command is not available
with segmented sweep
-221
Settings conflict; Marker->
function is not available in
zero span
-221
Settings conflict; Meas Type
was changed to Examine for
Exp Avg Mode.
-221
Settings conflict; Meas Type
was changed to Full for
Repeat Avg Mode
-221
Settings Conflict;
MinPts/RBW limit not met
-221
Settings conflict; Mixer
Type is not available when
in this band
-221
Settings conflict; Mkr -> CF
is not available when the
x-axis is time domain
-221
Settings conflict; MS gain
is not available in this
Mode
-221
Settings conflict; No meas
frequencies are above 3.6
GHz
-221
Settings conflict; Normal
detector is not allowed with
X scale is Log
-221
Settings conflict; Normalize
is not available when Scale
Type = Lin
-221
Settings conflict; Normalize
is not available while Demod
View is on
Keysight N9010A EXA Service Guide
If a Marker is a Fixed type marker, the marker's value does not
change from when it first became fixed. You cannot turn on or
change a Marker Function because there is no ongoing
measurement data to use for the marker function calculation.
MkrΔ->Span and MkrΔ->CF require that the selected marker be a
delta marker.
Most of the "Marker To" commands are not available with
segmented sweep turned on. So you cannot select these
functions.
Most of the "Marker To" functions are not available if you are in
zero span (span = 0 Hz, or time domain). So you cannot send the
commands for these functions.
Average Mode has been changed to Exponential. Full Meas Type is
not available for Exponential Average Mode therefore Meas Type
has been changed to Examine.
Average Mode has been changed to Repeat. Examine Meas Type is
not available for Repeat Average Mode therefore Meas Type has
been changed to Full.
The Mixer Type selection is not available in the K, E, W, F, D, G, Y, J
band because those bands do not allow Preselected Mixing.
The marker to center frequency functionality does not work when
the x-axis is in the time domain.
Mobile Station gain correction is not available in some Modes, or in
some measurements (for example, the SA measurement).
Optimize Preselector can only be performed on frequencies in high
band i.e. freqs above 3.6 GHz. The current setup does not have
input freqs (IF) in this range so an Optimize Preselector cannot be
performed.
The normal detector cannot be used when the x-axis scale is
logarithmic.
Why not? Are the results weird? Frequency slewed?
Normalize does not support Linear amplitude scale, since the
results are always presented as a dB ratio.
The normalization (correction) function cannot be used if you are
using the Demod View.